ESC 2026 removed HFmrEF: LVEF 41–49% is now HFrEF. Individual recommendations still carry their own
LVEF threshold (≤35%, ≤40%, <45%) — read the tag on each cell, not just the column.
Both phenotypes require symptoms and/or signs; an asymptomatic reduced EF is stage B (pre-HF) — see Prevention.
Echo test–retest variability is roughly ±5–8 EF points, so a value near a boundary is a range, not a threshold.
Device therapy ESC 2026 HF, with ESC 2021 Pacing / CRT and ESC 2022 Ventricular Arrhythmias where 2026 is silent
Therapy
Drug · start → target
HFrEF <50%
HFpEF ≥50%
ICD Primary and secondary prevention
Primary prevention: LVEF ≤35% despite ≥3 months of optimal FMT
NYHA class II–III
Expected survival >1 year with good functional status
Not within 40 days of MI
Reconsider at every generator replacement (residual risk falls as competing risks rise)
As it appears in the source
Recommendations
Classa
Levelb
An ICD is recommended in patients with symptomatic HFrEF (NYHA class II/III) of an ischaemic aetiology (unless they have had an MI in the prior 40 days), and an LVEF ≤35% despite ≥3 months of optimal FMT, provided they are expected to survive longer than 1 year with good functional status, to reduce the risk of sudden death and all-cause death.340,342,373,374
LVEF 36–49% (now HFrEF): no primary-prevention ICD recommendation — every primary-prevention threshold is ≤35%.ESC 2026 §6.2.1.2 describes the SCD-HeFT 23% reduction as “compared to amiodarone”; the primary publication reports it against placebo (amiodarone vs placebo HR 1.06).
was I A (ischaemic, LVEF ≤35%) · ESC 2021 HF §6.1.2
was no recommendation (HFmrEF 41–49%) · ESC 2021 HF §7.3.7
As it appears in the source
Recommendations
Classa
Levelb
An ICD should be considered in patients with symptomatic HFrEF (NYHA class II/III) of a non-ischaemic aetiology, and an LVEF ≤35% despite ≥3 months of optimal FMT, provided they are expected to survive longer than 1 year with good functional status, to reduce the risk of sudden death and all-cause death.340,341,344,375
RHR 0.87 (0.68-1.12), not significant — reduced sudden death but no overall mortality benefit
Benefit depends on competing non-sudden death; lower with age and comorbidity. Fibrosis (CMR) and genotype may refine risk (§6.2.1.3).
was IIa A (non-ischaemic, LVEF ≤35%) · ESC 2021 HF §6.1.2
As it appears in the source
Recommendations
Classa
Levelb
ICD therapy should be considered in patients with CAD, NYHA class I, and LVEF ≤30% despite ≥3 months of OMT.354
IIa
B
a Class of recommendation. b Level of evidence.
ischaemic aetiology, asymptomatic (NYHA I), LVEF ≤30% despite ≥3 months OMTESC 2022 Ventricular Arrhythmias §Table 24 · verified 2026-08-26Not addressed in ESC 2026 HF; ESC 2022 VA remains the source.
As it appears in the source
Recommendations
Classa
Levelb
An ICD is recommended in patients who have recovered from a ventricular arrhythmia causing haemodynamic instability (secondary prevention), and who are expected to survive for >1 year with good functional status, in the absence of reversible causes, or unless the ventricular arrhythmia has occurred <48 h after an MI, to reduce the risk of sudden death and all-cause death.346
I
A
a Class of recommendation. b Level of evidence.
secondary prevention — any LVEFESC 2026 HF §6.2.1.1 · Rec Table 6 · verified 2026-10-24Not restricted by LVEF. Excludes reversible causes and VA within 48 h of MI.
As it appears in the source
Recommendations
Classa
Levelb
ICD implantation for primary prevention is not recommended within 40 days of an MI as implantation during this period does not improve prognosis.358,359
III
B1
a Class of recommendation. b Level of evidence.
primary prevention within 40 days of MIESC 2026 HF §6.2.1.4 · Rec Table 6 · verified 2026-10-24
As it appears in the source
Recommendations
Classa
Levelb
ICD implantation is not recommended in patients in NYHA class IV with severe symptoms refractory to pharmacological therapy unless they are candidates for CRT, a ventricular assist device, or heart transplantation.374,385–387
III
C
a Class of recommendation. b Level of evidence.
NYHA IV refractory, not a CRT/LVAD/transplant candidateESC 2026 HF §6.2.1.3 · Rec Table 6 · verified 2026-10-24
As it appears in the source
Recommendations
Classa
Levelb
A wearable cardioverter-defibrillator may be considered in patients with HF who are at risk of sudden cardiac death for a limited period in order to increase survival as a bridge to decision for permanent ICD implantation or while listed for heart transplantation.381–384
Prior to generator replacement for CIEDs, a shared decision-making process between the patient and a cardiologist experienced in CIEDs and HF should be considered to ensure that the balance between risks and benefits still favour a CIED.376–380
IIa
C
a Class of recommendation. b Level of evidence.
shared decision before generator replacementESC 2026 HF §6.2.1.5 · Rec Table 6 · verified 2026-10-24>15% receive shocks after replacement despite no appropriate therapy from the first generator; residual SCD risk persists after LVEF improvement (§6.2.1.5).
no recommendationprimary preventionESC 2026 HF §6.2.1 · Rec Table 6 · verified 2026-10-24No primary-prevention recommendation. Secondary-prevention ICD (I A, HFrEF column) is not LVEF-restricted and applies here too.
was no recommendation (HFpEF) · ESC 2021 HF §8.4
Reassess LVEF before implant if FMT and CRT planning run in parallel
As it appears in the source
Recommendations
Classa
Levelb
CRT is recommended in patients with symptomatic HFrEF, LVEF ≤35%, despite optimal FMT, in SR with LBBB and QRS duration ≥150 ms in order to improve symptoms and reduce the risk of hospitalizations and death.388–398
R−20% (CRT-P and CRT-D); all-cause death reduced with CRT-D only
LVEF 36–49% in sinus rhythm (now HFrEF): no CRT recommendation unless there is a ventricular pacing indication for high-degree AV block.
was I A (LBBB ≥150 ms) · ESC 2021 HF §6.2
was no recommendation (HFmrEF 41–49%) · ESC 2021 HF §7.3.7
As it appears in the source
Recommendations
Classa
Levelb
CRT should be considered in patients with symptomatic HFrEF, LVEF ≤35%, despite optimal FMT, who are in SR with LBBB QRS morphology and QRS duration of 130–149 ms in order to improve symptoms and reduce morbidity and death.390,417
CRT should be considered in patients with symptomatic HFrEF, LVEF ≤35%, despite optimal FMT, who are in SR with non-LBBB QRS morphology and QRS duration ≥150 ms in order to improve symptoms and reduce morbidity and death.388–398
CRT may be considered in patients with symptomatic HFrEF, LVEF ≤35%, despite optimal FMT, who are in SR with non-LBBB QRS morphology and QRS duration of 130–149 ms in order to improve symptoms and reduce morbidity and death.391,397
CRT is not recommended in patients with a QRS duration <130 ms who do not have an indication for pacing due to high-degree AV block.410,424,425
III
A
a Class of recommendation. b Level of evidence.
QRS <130 ms without a high-degree AV block pacing indication — not recommendedESC 2026 HF §6.2.2.1 · Rec Table 7 · verified 2026-10-24
was III A (QRS <130 ms) · ESC 2021 Pacing §6.2
As it appears in the source
Recommendations
Classa
Levelb
Initiation of FMT and planning for CRT implantation may be considered simultaneously in patients with symptomatic HFrEF, an LBBB with QRS ≥150 ms, and LVEF ≤35%, to improve symptoms and reduce morbidity and death, although reassessment of LVEF should be conducted prior to CRT implantation.418–423
IIb
C
a Class of recommendation. b Level of evidence.
LVEF ≤35%LBBB ≥150 ms: start FMT and plan CRT in parallelESC 2026 HF §6.2.2 · Rec Table 7 · verified 2026-10-24For HF caused by abnormal conduction with a low likelihood of LVEF recovery; ideally implant before a first HFH (§6.2.2.3).
As it appears in the source
Recommendations
Classa
Levelb
CRT, rather than RV pacing, should be considered in patients with HFrEF regardless of NYHA class or QRS width who have an indication for ventricular pacing for high-degree AV block in order to reduce the risk of HFH and death.409
IIa
B1
a Class of recommendation. b Level of evidence.
LVEF <50%high-degree AV block needing ventricular pacing — CRT rather than RV pacing, any NYHA or QRSESC 2026 HF §6.2.2 · Rec Table 7 · verified 2026-10-24Applies across HFrEF, i.e. LVEF <50% — this is the only CRT indication covering LVEF 36–49%.
As it appears in the source
Recommendations
Classa
Levelb
An upgrade to CRT should be considered in patients with an LVEF ≤35% who have received a conventional pacemaker or an ICD and subsequently develop worsening HF despite optimal FMT and who have a significant proportion of RV pacing in order to reduce the risk of HFH or death.412
IIa
B1
a Class of recommendation. b Level of evidence.
LVEF ≤35%upgrade: pacemaker/ICD with significant RV pacing and worsening HFESC 2026 HF §6.2.2.2 · Rec Table 7 · verified 2026-10-24
CRT is recommended in patients with HFrEF.196,197,306,308
I
B
a Class of recommendation. b Level of evidence.
symptomatic AF with uncontrolled rate, AVJ ablation candidateESC 2021 Pacing §6.3.2 · verified 2026-08-26Not addressed in ESC 2026 HF, which lists CRT efficacy in AF as an evidence gap (§14).
As it appears in the source
Recommendations
Classa
Levelb
Patients with an LVEF ≤35% who have received a conventional pacemaker or an ICD and subsequently develop worsening HF despite OMT and who have a significant proportion of RV pacing should be considered for 'upgrade' to CRT.221
IIa
B
a Class of recommendation. b Level of evidence.
LVEF ≤35%, conventional pacemaker/ICD with significant RV pacing burden and new symptomatic HF despite OMT — CRT upgradeESC 2021 HF §6.2 · verified 2026-08-26ESC 2026 HF grades the comparable strategy (AV-node ablation + CRT, permanent AF, ≥1 HFH) IIa B1 — see next block.
As it appears in the source
Recommendations
Classa
Levelb
Atrioventricular node ablation combined with CRT should be considered in patients with severely symptomatic permanent AF and poor rate control despite medical therapy and at least one HFH to reduce symptoms, physical limitations, recurrent HFH, and death.672,673
IIa
B1
a Class of recommendation. b Level of evidence.
permanent AF, severe symptoms, poor rate control despite medical therapy, ≥1 HFH — AV-node ablation + CRTESC 2026 HF §9.1.1.4 · Rec Table 15 · verified 2026-10-24Not restricted by LVEF in the recommendation wording.
As it appears in the source
Recommendations
Classa
Levelb
CRT rather than standard RV pacing should be considered in patients with HFmrEF.
IIa
C
a Class of recommendation. b Level of evidence.
LVEF 41–49% (‘HFmrEF’ in ESC 2021 Pacing): AF with uncontrolled rate, AVJ ablation candidate — CRT preferred over RV pacingESC 2021 Pacing §6.3.2 · verified 2026-08-26
no recommendationsinus rhythmESC 2026 HF §6.2.2 · Rec Table 7 · verified 2026-10-24No CRT recommendation in sinus rhythm. ESC 2026 HF does not address pacing mode in HFpEF; ESC 2021 Pacing rows below still apply.
was no recommendation (HFpEF) · ESC 2021 HF §8.4
As it appears in the source
Recommendations
Classa
Levelb
RV pacing should be considered in patients with HFpEF.188,196,323
IIa
B
a Class of recommendation. b Level of evidence.
symptomatic AF with uncontrolled rate, AVJ ablation candidate — RV pacing preferred over CRTESC 2021 Pacing §6.3.2 · verified 2026-08-26
As it appears in the source
Recommendations
Classa
Levelb
CRT may be considered in patients with HFpEF.
IIb
C
a Class of recommendation. b Level of evidence.
symptomatic AF with uncontrolled rate, AVJ ablation candidate — CRT may be consideredESC 2021 Pacing §6.3.2 · verified 2026-08-26
RV-pacing burden review Any CIED
De novo or worsening HFrEF in a patient with a pacemaker, ICD or CRT
As it appears in the source
Recommendations
Classa
Levelb
Review of the proportion of RV pacing should be considered in all patients with CIED and de novo or worsening HFrEF.
no recommendationESC 2026 HF §6.2.3 · verified 2026-10-24“no recommendations can currently be made”No RCT with patient-centred outcomes on efficacy or long-term safety of CSP in HFrEF. LBBAP may be a feasible alternative when CRT implantation is challenging (§6.2.3). See EHRA 2023 consensus for implantation technique.
no recommendationESC 2026 HF §6.2.3 · verified 2026-10-24Not addressed.
Interatrial shunt
no recommendationESC 2026 HF §6.2.6 · verified 2026-10-24No recommendation. RELIEVE-HF (any LVEF) neutral overall — fewer events with reduced LVEF, more with HFpEF; REDUCE LAP-HF II (LVEF ≥40%) neutral (§6.2.6).
ClassI is recommendedIIa should be consideredIIb may be consideredIII not recommended / harmnone no recommendation
LevelA multiple RCTs or meta-analysisB1 ≥1 adequately powered RCT, or meta-analysis of suchB2 ≥2 well-controlled non-randomised studies, or meta-analysis of small RCTsC preliminary evidence or expert consensus
ESC 2026 split Level B into B1/B2 and regraded every recommendation. Older documents (ESC 2021–2025) still show
B (single RCT or large non-randomised studies) — a B is not a lower grade than a B1, it is the previous system.
A LVEF ≤35% tag is the recommendation's own threshold; a grey “was …” line gives the grade it replaced and where that came from.